Turning on the Sensor Feature
DEVICE: 523/723 (Revel), 522/722, Guardian
Important The sensor must be turned on, started and initialized to report glucose measurements.
- Go to the EDIT SETTINGS screen.
- For 523/723 (Revel)
- Main Menu > Sensor > Edit Settings
- For 522/722 insulin pump models and Guardian
- Main Menu > Sensor > Sensor Setup > Edit Settings
- The EDIT SETTINGS screen appears with Sensor: Off selected.
- Press ACT. The SENSOR ON/OFF screen appears.
- Select On, then press ACT. The EDIT SETTINGS screen appears showing the sensor is now turned on.
Entering Transmitter ID
DEVICE: 523/723 (Revel), 522/722, Guardian
Important You must link your transmitter ID to your insulin pump or CGM monitor the first time you use your transmitter. Each time you replace your insulin pump, CGM monitor or transmitter, you will need to input a transmitter ID again to enable communication.
- Make sure the SENSOR feature is turned on.
- For 523/723 (Revel)
- Main Menu > Sensor > Edit Settings > Sensor: On
- For 522/722 insulin pump models and Guardian
- Main Menu > Sensor > Sensor Setup > Edit Settings > Sensor: On
- Make sure the EDIT SETTINGS screen is open.
- For 523/723 (Revel)
- Main Menu > Sensor > Edit Settings
- For 522/722 insulin pump models and Guardian
- Main Menu > Sensor > Sensor Setup > Edit Settings
- Select Transmtr ID (Transmitter ID for Guardian), then press ACT. The SET TRANSMITTER ID screen shows 7 dashes for the Transmitter ID, the first of which (on the left) is flashing.The 7-digit ID (or serial number) is located on the flat side of the transmitter.
- Select the first number of the Transmitter ID (the number on the screen must match the first number of the Transmitter ID).
- Press ACT. The second dash flashes.
- Select the next number, then press ACT.
- Repeat step 6 until all 7 digits of the Transmitter ID have been entered.
- A message shows that the Transmitter ID has been changed. After about 30 seconds, the EDIT SETTINGS screen appears with the Transmitter ID you entered.